This hemocytometer calculator turns square counts into cell concentration. Enter the live cells counted in each large square, the dead (blue) cells if you used trypan blue, and the dilution, and it gives cells per mL, viability and the volume to pipette for the number of cells you want to seed, with warnings when the count is too low or the squares disagree. Paste a list of samples to calculate them all. It runs in your browser.
| Mean count | Cells/mL |
|---|---|
| 10 | 2.0 × 10⁵ |
| 25 | 5.0 × 10⁵ |
| 50 | 1.0 × 10⁶ |
| 100 | 2.0 × 10⁶ |
Average the count per large square, multiply by the dilution and by 10,000. An average of 50 with a 1:1 trypan blue dilution is 50 × 2 × 10⁴ = 1 × 10⁶ cells/mL.
At least four large squares and 100+ cells in total for a reliable estimate.
Live ÷ (live + dead) × 100, counting blue cells as dead.
Count cells touching the top and left borders and ignore those on the bottom and right.
No. Everything runs in your browser.
Cells per mL = mean live count per large square × dilution factor × 10⁴, because each 1 mm × 1 mm square of a Neubauer chamber holds 0.1 µL (the square volume is editable). Viability = live ÷ (live + dead) from trypan blue counts. Seeding volume = cells needed ÷ concentration. The coefficient of variation between squares and a total count under 100 are flagged. Batch mode processes a list of samples.